- Crashes may occur when reading in a new XEASY spectrum.
- Try to open the spectrum via Tools->Open Spectrum. This will open a MonoScope window with your spectrum. If you can work normally in this window, the error resides in the Spectrum Type definition. Either you chose a definition not suited to your spectrum or your definitions are faulty.
- If normal work is not possible via Tools->Open Spectrum, have a look at the .param file of the spectrum, especially if it was produced via uxnmr2xeasy. An actual example produced by uxnmr2xeasy is shown below. The example will crash immediately, and even if the first mistake is fixed, Show folded will not work, since CARA needs to know the folding type which is not correctly defined in the example.
Spectrometer frequency in w3 .. 750.130005 Spectral sweep width in w1 .... 39.626991 Spectral sweep width in w2 .... 31.557470 Spectral sweep width in w3 .... 8.416050 Maximum chemical shift in w1 .. 199713.796875 <-- Should be around 12 ppm !!! Maximum chemical shift in w2 .. 199630.796875 Maximum chemical shift in w3 .. 199566.093750 Size of spectrum in w1 ........ 128 Size of spectrum in w2 ........ 128 Size of spectrum in w3 ........ 1024 Submatrix size in w1 .......... 16 Submatrix size in w2 .......... 16 Submatrix size in w3 .......... 128 Permutation for w1 ............ 3 Permutation for w2 ............ 2 Permutation for w3 ............ 1 Folding in w1 ................. NO <------------- Should be RSH, ... !!! Folding in w2 ................. NO Folding in w3 ................. NO Type of spectrum .............. ? Identifier for dimension w1 ... N Identifier for dimension w2 ... C Identifier for dimension w3 ... H
- Trouble-shooting: Try to open another spectrum of the same type. If it also fails, this indicates problems with SpectrumType definition. Try to open the same spectrum after loading the most recent Cara template from the TemplatesPage. Try it on another computer or even another operating system.
Return to CARAcrashes